1 The Test System Interface Daemon (TSID) will provide a connection-oriented | |
2 (SOCK_STREAM) UNIX domain local socket at /tmp/fc_rftest_socket, and will speak | |
3 an ASCII line-based command-response protocol on this socket. All command and | |
4 response lines will end with \n only (no \r), and each TSID response including | |
5 the initial greeting on connection establishment will begin with a '+' or '-' | |
6 character to indicate OK or error, respectively, like in POP3. | |
7 | |
8 The following commands are currently defined: | |
9 | |
10 vcxo-cal-setup band arfcn | |
11 | |
12 This command is sent to the TSID to prepare the RF test system for the VCXO | |
13 calibration procedure. The first argument after the command keyword is the | |
14 band to be used (850, 900, 1800 or 1900) and the second argument is the ARFCN | |
15 in that band on which the DUT will be transmitting. | |
16 | |
17 freq-meas hint | |
18 | |
19 This command directs the RF test system to take a frequency offset measurement | |
20 (vcxo-cal-setup must have been performed previously) and report it back to the | |
21 client in the response line. The hint argument after the freq-meas keyword will | |
22 be sent by the VCXO calibration program (fc-rfcal-vcxo) to tell the RF test | |
23 system what kind of frequency offset measurement is being made, i.e., which step | |
24 in the VCXO calibration sequence is being performed; the hint keywords remain | |
25 to be defined. | |
26 | |
27 signal-gen-setup band | |
28 | |
29 This command is sent to the TSID to prepare the RF test system for tests in | |
30 which the latter acts as a signal generator. The argument after the command | |
31 keyword is the band number (850, 900, 1800 or 1900), and it selects the | |
32 downlink frequency band in which the signal generator will need to operate. | |
33 | |
34 signal-gen-sine arfcn offset level | |
35 | |
36 This command directs the RF test system to turn on its signal generator and put | |
37 out a pure sine wave at the frequency corresponding to the specified ARFCN (in | |
38 the downlink band previously selected with signal-get-setup) plus the specified | |
39 offset in kHz (-67 or +67 with TI's Rx path calibration procedures), and at the | |
40 specified Tx power level in dBm. | |
41 | |
42 signal-gen-off | |
43 | |
44 Turn off the signal generator. | |
45 | |
46 The commands for Tx power level calibration remain to be defined. |