comparison doc/Rx-cal-theory @ 65:e7b9a9903f14

doc/Rx-cal-theory: typo fix
author Mychaela Falconia <falcon@freecalypso.org>
date Sun, 28 May 2017 05:10:32 +0000
parents 131abadbd74d
children
comparison
equal deleted inserted replaced
64:65577bb967f7 65:e7b9a9903f14
66 66
67 Note that the LoCosto document indicates the use of a different signal generator 67 Note that the LoCosto document indicates the use of a different signal generator
68 level (-68.5 dBm instead of -74.5), and tells the operator to set the AGC gain 68 level (-68.5 dBm instead of -74.5), and tells the operator to set the AGC gain
69 to 0 dB. The latter must be a LoCosto-ism, as it is definitely not correct for 69 to 0 dB. The latter must be a LoCosto-ism, as it is definitely not correct for
70 our Calypso+Iota+Rita chipset. If we set the signal generator level to 70 our Calypso+Iota+Rita chipset. If we set the signal generator level to
71 -68.5 dBm, we would need to get the Rita AGC to 28 dB, but the Mother of 71 -68.5 dBm, we would need to set the Rita AGC to 28 dB, but the Mother of
72 FreeCalypso decided to go with the safer choice of using the Sara settings of 72 FreeCalypso decided to go with the safer choice of using the Sara settings of
73 -74.5 dBm generator level, 34 dB AGC. 73 -74.5 dBm generator level, 34 dB AGC.
74 74
75 FreeCalypso Rx calibration tools 75 FreeCalypso Rx calibration tools
76 ================================ 76 ================================