FreeCalypso > hg > freecalypso-docs
comparison CMU200-maintenance-notes @ 85:7fc66c5b54dd
CMU200-maintenance-notes: updates from Michael VK2BEA
author | Mychaela Falconia <falcon@freecalypso.org> |
---|---|
date | Fri, 21 Jan 2022 05:26:17 +0000 |
parents | d2fef140ed53 |
children | 2ac06a49dfbc |
comparison
equal
deleted
inserted
replaced
84:d2fef140ed53 | 85:7fc66c5b54dd |
---|---|
26 ======================= | 26 ======================= |
27 | 27 |
28 This board is common among CMU200, CMU300 and CRTU-RU instruments from R&S - or | 28 This board is common among CMU200, CMU300 and CRTU-RU instruments from R&S - or |
29 at least these are the ones I know - maybe there are others I don't know about. | 29 at least these are the ones I know - maybe there are others I don't know about. |
30 This board encapsulates the instrument's main RF Rx and Tx chains: on the Rx | 30 This board encapsulates the instrument's main RF Rx and Tx chains: on the Rx |
31 side it takes RF input from the front end and performs triple IF downconversion | 31 side it takes RF input from the front end and performs triple (or quadruple, |
32 to 10.7 MHz IF3, and on the Tx side it takes 13.85 MHz IF3 input and upconverts | 32 explained later in this article) IF downconversion to 10.7 MHz IF3, and on the |
33 it to RF output, going through IF2 and IF1 in the process - triple IF in both | 33 Tx side it takes 13.85 MHz IF3 input and upconverts it to RF output, going |
34 directions. | 34 through IF2 and IF1 in the process - triple or quadruple IF in both directions, |
35 as explained in more detail later in this article. | |
35 | 36 |
36 Every CMU200 instrument always has one RXTX board - it is an absolutely required | 37 Every CMU200 instrument always has one RXTX board - it is an absolutely required |
37 component irrespective of option configurations. The hardware architecture of | 38 component irrespective of option configurations. The hardware architecture of |
38 this instrument also has a place for an optional second RXTX board, providing a | 39 this instrument also has a place for an optional second RXTX board, providing a |
39 complete second Rx and Tx channel - however, as far as I can tell, CMU200 | 40 complete second Rx and Tx channel - however, as far as I can tell, CMU200 |
66 the two IF1 Rx paths fails, then only that frequency range will be affected, | 67 the two IF1 Rx paths fails, then only that frequency range will be affected, |
67 or if both fail, the observed loss will typically be different between the two | 68 or if both fail, the observed loss will typically be different between the two |
68 frequency ranges. The failure symptom is unexpected large attenuation: | 69 frequency ranges. The failure symptom is unexpected large attenuation: |
69 sometimes around 5 to 6 dB of loss, othertimes as much as 25 dB of loss. | 70 sometimes around 5 to 6 dB of loss, othertimes as much as 25 dB of loss. |
70 | 71 |
72 The internal loopback test invoked from the Maintenance menu is a good first | |
73 step in diagnosis. In this test the instrument software configures both Rx and | |
74 Tx chains to connect to RF1 (and then RF2 if you press Continue), and then it | |
75 tests a longish list of different frequencies in sequence, spanning the full | |
76 range from 10 to 2700 MHz. For each test frequency, the instrument software | |
77 configures both the signal generator and the Rx chain, and it reports what was | |
78 measured on Rx vs. what was put out on Tx. The test is considered a failure if | |
79 nothing was received or if the Rx signal level was too far from the expected | |
80 value, otherwise the test is declared as passed. | |
81 | |
82 If the loopback test fails at every frequency with no signal detected, then you | |
83 don't really know what's going on, and you will need to manually test Rx and Tx | |
84 separately (using an external spectrum analyzer and an external signal | |
85 generator) in order to figure out what is broken. However, out of the commonly | |
86 observed failure modes, dead Tx will produce this symptom. | |
87 | |
88 If the Tx side is OK but Rx IF1 filters (one or both paths) have gone bad, the | |
89 visible symptom in the loopback test will be Rx signal level that is lower than | |
90 it should be. The instrument software may declare the test as either passed or | |
91 failed depending on the magnitude of the error: in this Mother's experience, | |
92 when one of the two IF1 paths on my CMU200 developed a loss of some 5.8 dB, the | |
93 loopback test was reported as passing - but a closer look at the numbers in the | |
94 report window showed the unexpected attenuation. | |
95 | |
96 Because different input frequency ranges are handled via different Rx paths as | |
97 explained in the following section, when Rx IF1 filters fail, the loss behaviour | |
98 will be frequency-dependent. In the internal loopback test, you will see one | |
99 behaviour for frequencies from 10 to 1200 MHz, then a marked change for | |
100 frequencies from 1205 to 2200 MHz, and then another change (most likely a | |
101 reversion to low frequency behaviour) at the highest frequencies above 2200 MHz. | |
102 | |
103 We do not currently know if there are any other failure modes elsewhere in the | |
104 CMU200 instrument that can also cause a stepwise change in behaviour at these | |
105 frequency cutover points. It is my (Mother Mychaela's) suspicion that the front | |
106 end may have some filters too, each covering a wide frequency swath, with | |
107 instrument software switching these filters depending on the configured | |
108 listening frequency - but we don't know for certain if any such additional | |
109 filters are there or not. If you find yourself wondering whether the problem | |
110 you are seeing is in the RXTX board or the front end, the best way to narrow it | |
111 down would be to remove the semi-rigid coax pieces that carry RF between the two | |
112 and use an external spectrum analyser to look at the Tx output from the RXTX | |
113 board and/or the Rx output from the front end. | |
114 | |
71 RXTX board architecture explained | 115 RXTX board architecture explained |
72 ================================= | 116 ================================= |
73 | 117 |
74 Unfortunately R&S' official service manual for CMU200 instruments is only a part | 118 Unfortunately R&S' official service manual for CMU200 instruments is only a part |
75 swapper guide: it tells you which boards do what in general terms and tells you | 119 swapper guide: it tells you which boards do what in general terms and tells you |
76 how to remove and replace each part, but no schematics, and no detailed | 120 how to remove and replace each part, but no schematics, and no detailed |
77 explanation of what happens inside each board. However, I draw the reader to | 121 explanation of what happens inside each board. They do provide a little bit of |
78 the block diagram on page 3.2 of this manual - this block diagram does provide | 122 info: I draw the reader to the block diagram on page 3.2 of this manual - this |
79 an important starting point for understanding what happens inside the RXTX | 123 block diagram does provide an important starting point for understanding what |
80 board. | 124 happens inside the RXTX board - however, it is simplified and incomplete. |
81 | 125 |
82 In the Tx direction, 13.85 MHz IF3 comes in from the digital board - or from | 126 In the Tx direction, 13.85 MHz IF3 comes in from the digital board - or from |
83 B68 board in WCDMA test modes. This Tx IF3 is mixed with Tx LO3 to produce | 127 B68 board in WCDMA test modes. This Tx IF3 is mixed with Tx LO3 to produce |
84 Tx IF2. This Tx IF2 is fixed at 487.52 MHz, thus one would think that Tx LO3 | 128 Tx IF2. This Tx IF2 is fixed at 487.52 MHz, thus one would think that Tx LO3 |
85 frequency ought to be fixed as well - but it seems to be a synthesized variable | 129 frequency ought to be fixed as well - but it seems to be a synthesized variable |
86 frequency. (Remember, all of this understanding is from reverse engineering, | 130 frequency, and the manual describes it as "LO3TX with small tuning range". |
87 hence we can only figure out so much.) | 131 Calculations done by Michael VK2BEA put Tx LO3 at 473.67 MHz (needs |
132 confirmation), but it is still not clear why it is a synthesized frequency | |
133 "with small tuning range", as opposed to simply fixed. | |
88 | 134 |
89 Tx IF2 of 487.52 MHz is then passed through a pair of identical SAW filters, | 135 Tx IF2 of 487.52 MHz is then passed through a pair of identical SAW filters, |
90 Sawtek 855272 - two cascaded identical filters, with an amplifier in between. | 136 Sawtek 855272 - two cascaded identical filters, with an amplifier in between. |
91 This SAW filter has a center frequency of 479.75 MHz with 20 MHz bandwidth, | 137 This SAW filter has a center frequency of 479.75 MHz with 20 MHz bandwidth, |
92 thus the passband spans from 469.75 to 489.75 MHz. Notice how Tx IF2 of | 138 thus the passband spans from 469.75 to 489.75 MHz. Notice how Tx IF2 of |
93 487.52 MHz stands just 2.23 MHz away from the edge of the passband - is it | 139 487.52 MHz stands just 2.23 MHz away from the edge of the passband - is it |
94 intentional? What are they filtering? Without original design notes, we can | 140 intentional? What are they filtering? Without original design notes, we can |
95 only guess. As I shall explain later in this article, one of these two Tx IF2 | 141 only guess. As I shall explain later in this article, one of these two Tx IF2 |
96 SAW filters is a component prone to failure. | 142 SAW filters is a component prone to failure. |
143 | |
144 [Note from Michael VK2BEA: "The LO frequency is only 13.85 MHz from the IF. It | |
145 makes sense to shift this to the edge of the passband to help the suppression | |
146 of LO feed through. Also explains the use of SAW filters (sharp skirts) and | |
147 that there are two."] | |
97 | 148 |
98 After these cascaded SAW filters, Tx IF2 is mixed with LO2. Unlike LO1 and LO3, | 149 After these cascaded SAW filters, Tx IF2 is mixed with LO2. Unlike LO1 and LO3, |
99 there is only one LO2 for both Rx and Tx, and it is fixed at 1329.6 MHz. When | 150 there is only one LO2 for both Rx and Tx, and it is fixed at 1329.6 MHz. When |
100 Tx IF2 at fixed 487.52 MHz is mixed with LO2 at fixed 1329.6 MHz, the output of | 151 Tx IF2 at fixed 487.52 MHz is mixed with LO2 at fixed 1329.6 MHz, the output of |
101 this mixer will always contain two frequencies: 842.08 MHz and 1817.12 MHz. | 152 this mixer will always contain two frequencies: 842.08 MHz and 1817.12 MHz. |
105 high Tx IF1, controlling switches before and/or after the filters. I have not | 156 high Tx IF1, controlling switches before and/or after the filters. I have not |
106 investigated to see if the frequency ranges for high vs. low Tx IF1 are the same | 157 investigated to see if the frequency ranges for high vs. low Tx IF1 are the same |
107 as on the Rx side or not - maybe they are the same, maybe they are different. | 158 as on the Rx side or not - maybe they are the same, maybe they are different. |
108 | 159 |
109 After Tx IF1 output is combined or switched from the two filters, it is mixed | 160 After Tx IF1 output is combined or switched from the two filters, it is mixed |
110 with Tx LO1 to produce the final RF output. The mixer that does this job is | 161 with Tx LO1 to produce an output that may or may not be final RF. The mixer |
111 MACOM SM4T, which is one of the larger, prominently visible components on the | 162 that does this job is MACOM SM4T, which is one of the larger, prominently |
112 board. There also seems to be a fourth mixer and LO stage that kicks in only | 163 visible components on the board. Tx LO1 has "large tuning range and very fine |
113 for frequencies above 2200 MHz, but I haven't really studied that one as my main | 164 frequency resolution used for setting the desired transmitter frequency" - quote |
114 interest is in the classic cellular frequency bands, 1900 MHz and below. | 165 from the manual; by doing some frequency arithmetics, we can see that this Tx |
166 LO1 tuning range needs to span from 1827.12 to 3042.08 MHz in order to produce | |
167 output frequencies from 10 to 2200 MHz starting from 842.08 MHz or 1817.12 MHz | |
168 IF1. (LO1 - IF1 is the desired output frequency, whereas the sum will be a | |
169 much higher frequency above 2.7 GHz - I presume that the latter must be | |
170 suppressed by some LPF somewhere.) | |
171 | |
172 The "RF" output from Tx SM4T mixer (LO1-IF1 as explained above) is indeed the | |
173 final RF output going to the front end for output frequencies below 2200 MHz. | |
174 In the uppermost frequency range of 2200 to 2700 MHz, a fourth mixer and LO | |
175 stage come into play - NOT shown on the block diagram in the manual! In this | |
176 highest frequency range, the output from SM4T mixer should be considered a | |
177 fourth IF - but because it is not covered at all in the manual and not named, | |
178 we have to invent our own name for it. I (Mother Mychaela) propose that we | |
179 call it IF0, and refer to the corresponding LO as LO0 - this way we remain | |
180 consistent with official naming that puts IF1 closest to RF and IF3 closest to | |
181 digital. | |
182 | |
183 The preliminary analysis by Michael VK2BEA is that Tx LO0 frequency is fixed at | |
184 3318.46 MHz (same as its counterpart on the Rx side), with IF0 (taking the place | |
185 of lower RF) ranging from 1118.46 to 618.46 MHz (reverse range) to produce final | |
186 output frequencies of 2200 to 2700 MHz. However, these numbers have NOT been | |
187 confirmed by actual measurements yet. | |
115 | 188 |
116 On the Rx side the same process happens in reverse, but the specific frequencies | 189 On the Rx side the same process happens in reverse, but the specific frequencies |
117 used for IF1, IF2 and IF3 are slightly different. At first there is a stage | 190 used for IF1, IF2 and IF3 are slightly different. At first there is a stage |
118 that only kicks in for frequencies above 2200 MHz (bypassed otherwise), and | 191 that only kicks in for frequencies above 2200 MHz (bypassed otherwise), and |
119 then there is an SM4T mixer (identical to the one on Tx side) that takes in RF | 192 then there is an SM4T mixer (identical to the one on Tx side) that takes in RF |
158 cascaded Sawtek 855272 filters, two identical filters with an amplifier in | 231 cascaded Sawtek 855272 filters, two identical filters with an amplifier in |
159 between, exactly the same as on the Tx side. Then there is Rx LO3 and the final | 232 between, exactly the same as on the Tx side. Then there is Rx LO3 and the final |
160 mixer, producing Rx IF3 at 10.7 MHz that goes to the digital board, to the rear | 233 mixer, producing Rx IF3 at 10.7 MHz that goes to the digital board, to the rear |
161 panel BNC output and to the WCDMA board (B68) if the latter is present. | 234 panel BNC output and to the WCDMA board (B68) if the latter is present. |
162 | 235 |
236 Frequency conversion tables | |
237 =========================== | |
238 | |
239 Michael VK2BEA worked out a pair of frequency conversion tables, one for Rx and | |
240 one for Tx. Here are these tables, with further corrections by Mother Mychaela: | |
241 | |
242 Rx frequency conversion, RF to IF1: | |
243 | |
244 RF (MHz) LO0 (MHz) IF0 (MHz) LO1 (MHz) IF1 (MHz) | |
245 ---------------------------------------------------------------------- | |
246 10-1200 1826.115-3016.115 1816.115 | |
247 1200-2200 2043.085-3043.085 843.085 | |
248 2200-2700 3318.46 1118.46-618.46 2934.575-2434.575 1816.115 | |
249 | |
250 Rx frequency conversion, IF1 to IF3: | |
251 | |
252 IF1 (MHz) LO2 (MHz) IF2 (MHz) LO3 (MHz) IF3 (MHz) | |
253 -------------------------------------------------------- | |
254 1816.115 1329.6 486.515 497.215 10.7 | |
255 843.085 1329.6 486.515 497.215 10.7 | |
256 | |
257 Tx frequency conversion, IF3 to IF1: | |
258 | |
259 IF3 (MHz) LO3 (MHz) IF2 (MHz) LO2 (MHz) IF1 (MHz) | |
260 --------------------------------------------------------- | |
261 13.85 473.67 487.52 1329.6 1817.12 | |
262 13.85 473.67 487.52 1329.6 842.08 | |
263 | |
264 Tx frequency conversion, IF1 to RF: | |
265 | |
266 IF1 (MHz) LO1 (MHz) IF0 (MHz) LO0 (MHz) RF(MHz) | |
267 ------------------------------------------------------------------ | |
268 1817.12 1827.12-3017.12 10-1200 | |
269 842.08 2042.08-3042.08 1200-2200 | |
270 1817.12 2935.58-2435.58 1118.46-618.46 3318.46 2200-2700 | |
271 | |
272 Notes: | |
273 | |
274 * In Michael's original version each table covered the full chain from RF on | |
275 one end to IF3 on the other end, but I (Mychaela) had to split each table | |
276 into two in order to fit within 80 columns. | |
277 | |
278 * The numbers for LO3 (473.67 MHz for Tx, 497.215 MHz for Rx) are from Michael; | |
279 I (Mychaela) have not verified them. | |
280 | |
281 * All details for the IF0/LO0 stage (upper frequency range) are from Michael; | |
282 his notes indicate that the numbers are confirmed for Rx, but not for Tx. | |
283 | |
163 How these RXTX boards fail | 284 How these RXTX boards fail |
164 ========================== | 285 ========================== |
165 | 286 |
166 There are 3 specific components on this RXTX board that have been seen to fail | 287 There are 3 specific components on this RXTX board that have been seen to fail |
167 over and over in the field: | 288 over and over in the field: |
187 The complete RXTX board assembly consists of the traditional PCBA plus heavy | 308 The complete RXTX board assembly consists of the traditional PCBA plus heavy |
188 metal shields on both sides; the front and back metal shield pieces are custom- | 309 metal shields on both sides; the front and back metal shield pieces are custom- |
189 made for this board, with individually shielded cavities matching different | 310 made for this board, with individually shielded cavities matching different |
190 sections of the board. Some (not all) of these cavities are filled with a | 311 sections of the board. Some (not all) of these cavities are filled with a |
191 special black foam called Eccosorb - it is an RF absorber, presumably added to | 312 special black foam called Eccosorb - it is an RF absorber, presumably added to |
192 prevent these cavities from acting as parasitic oscillators. Trouble occurs | 313 lower the Q of these cavities to prevent parasitic oscillations. Trouble occurs |
193 when this Eccosorb foam comes into direct contact with metal surfaces of | 314 when this Eccosorb foam comes into direct contact with metal surfaces of |
194 components on the board: the result is galvanic corrosion, a process that takes | 315 components on the board: the result is galvanic corrosion, a process that takes |
195 many years before it results in component failure. The reason why only 3 | 316 many years before it results in component failure. The reason why only 3 |
196 particular filter components fail is because they got the bad luck of residing | 317 particular filter components fail is because they got the bad luck of residing |
197 in cavities with Eccosorb - the other identical components that don't fail | 318 in cavities with Eccosorb - the other identical components that don't fail |
198 reside in cavities without Eccosorb. | 319 reside in cavities without Eccosorb. |
199 | 320 |
200 We don't know how R&S allowed this design flaw to escape and remain in their | 321 [Note from Michael VK2BEA: "The copper surface of these filters form an integral |
201 sold and field-deployed products: there is the "innocent" explanation that they | 322 part of the component. It is this copper that forms the cavity of the combline |
202 simply didn't notice, and there is the conspiratorial view that this slow | 323 filter. When this is compromised by corrosion, the filter is detuned and there |
203 failure mechanism is intentional as in planned obsolescense - pick your choice | 324 is leakage causing excessive loss."] |
204 of hypothesis. | 325 |
326 It appears that R&S only noticed this design flaw toward the end of "product | |
327 life" of these instruments, probably because failures occur only after many | |
328 years. Some of the newer boards have had modifications to prevent contact | |
329 between Eccosorb and the two troubled Rx filters, either by way of thinner | |
330 Eccosorb fill or by way of an added plastic barrier. It is not clear if these | |
331 modifications were applied to newer produced RXTX boards from the start, or if | |
332 they are a result of field service repairs. | |
205 | 333 |
206 How to repair failed boards | 334 How to repair failed boards |
207 =========================== | 335 =========================== |
208 | 336 |
209 All 3 of the failing filter components (one SAW filter part and two ceramic | 337 All 3 of the failing filter components (one SAW filter part and two ceramic |
337 and then covering the component with Kapton tape before putting the board back | 465 and then covering the component with Kapton tape before putting the board back |
338 into its metal casing. | 466 into its metal casing. |
339 | 467 |
340 Unlike the failing filters, this MACOM SM4T mixer is still available new - but | 468 Unlike the failing filters, this MACOM SM4T mixer is still available new - but |
341 it's an expensive component, so let's protect these mixers from corrosion. | 469 it's an expensive component, so let's protect these mixers from corrosion. |
470 | |
471 [Michael VK2BEA notes: "The case of the mixer is purely for shielding and is | |
472 much thicker than the thin copper of the filter that is essential for | |
473 operation." Mother Mychaela's response: it may be so, but if you are going to | |
474 take the RXTX board out of your CMU, take it out of its metal casing and either | |
475 replace filters or at least cover them with Kapton tape for protection, it | |
476 won't hurt to put the same Kaptop tape on the mixers too - and the signs of | |
477 corrosion are very real.] |